Einfach Bitcoin kaufen: Mit dem Code "FINANZEN" sparen Sie 21% der Gebühren für 6 Monate bei Coinfinity. Jetzt loslegen -w-
06.05.2014 15:15:03

Alere Voluntarily Recalls PT/INR Professional Test Strips In The U.S

(RTTNews) - Alere Inc. (IMA , ALR) Tuesday announced a voluntary Class 1 recall of the Alere INRatio2 PT/INR Professional Test Strips (PN 99008G2) in the U.S., as a result of nine serious adverse event reports, out of which three reports involved bleeding associated with patient deaths.

The reason for the adverse event reports was significantly different test results between the Alere INRatio2 PT/INR Professional Test Strip and the local laboratory plasma INR test. The Alere INRatio2 PT/INR Professional Test Strip results were between 3.1 - 12.2 INR units lower than the laboratory result.

Alere said it will transition its customers from the current Alere INRatio2 PT/INR Professional Test Strip to the Alere INRatio PT/INR Test Strip (PN 100139).

The voluntary recall does not include the Alere INRatio PT/INR Test Strip (PN 100071), which is utilized by patient self-testers for home INR monitoring.

The company said that the root cause for this issue has not yet been determined; therefore it cannot determine the patient conditions or circumstances that may contribute to the discrepancy.

Nachrichten zu Alere Incmehr Nachrichten

Keine Nachrichten verfügbar.

Analysen zu Alere Incmehr Analysen

Eintrag hinzufügen
Hinweis: Sie möchten dieses Wertpapier günstig handeln? Sparen Sie sich unnötige Gebühren! Bei finanzen.net Brokerage handeln Sie Ihre Wertpapiere für nur 5 Euro Orderprovision* pro Trade? Hier informieren!
Es ist ein Fehler aufgetreten!